Development and implementation of a wireless real-time radiation monitoring system for LINAC beam delivery monitoring

Các tác giả

  • Vo Hong Hai
  • Phuc Tri Toan Nguyen
  • Hieu Trung Nguyen

Từ khóa

LINAC, beam delivery, embedded electronic, real-time monitoring, GM counter

DOI:

https://doi.org/10.32508/stdj.v27i2.4303

Tóm tắt

Introduction: In this work, we developed a wireless, remote, real-time radiation monitoring system designed to oversee beam delivery in a radiation therapy room equipped with a medical linear accelerator (LINAC).

Methods: This system utilizes a Geiger-Müller detector paired with embedded electronic hardware to accurately record radiation count rates in real-time. The data collected by this system are transmitted through LAN/WAN networks to the internet, ensuring instantaneous accessibility. In addition, a web server and mobile application were developed to display, receive, and archive the data from the radiation counter.

Results: Our system was deployed in the radiation therapy room of the Oncology Hospital in Ho Chi Minh City, demonstrating a remarkable data reception rate of up to 99.8% over a three-day test period from June 27th to 29th, 2022. The system effectively identified beam-on instances and provided precise measurements of the number and duration of beam-on events.

Conclusion: This study demonstrates the feasibility of remote real-time radiation monitoring in medical settings and highlights the potential for enhancing radiation safety and treatment efficacy in external beam radiotherapy.

Lượt tải

Chưa có dữ liệu tải xuống.

Lượt tải xuống

Đã Xuất bản

2024-06-30