Simulating the operation of 10 MHz quartz crystal microbalance under the effection of fluid used couple field method
Abstract
This article presents the simulation of operation of 10 MHz resonance frequency quartz crystal microblance device QCM in case of fluid domain (blood, water, high viscosity fluid …). Because QCM device operates in piezoelectric substrate, simulation has to use the structure-electric couple field element (SOLID 226). Determination of behaviors of device in the case of fluid domain is important for applications. Couple field method (structure-fluid) including finite element method and computational fluid dynamic method was used to determine output signals. The loading quantitative method on surface’s devices was built for the applications in biological and chemical sensors.
điểm /
đánh giá
Published
2017-02-10
Issue
Section
ARTILES
Copyright belongs to VNU-HCM “Science and Technology Development” Journal. Any copy or reprinting of any form must be permitted by the Journal.